NeuroTrac® Myoplus PRO
Clinical device delivering EMG, ETS and NMES. Has built in games for engagement with EMG function.
Please note we cannot supply this product to US addresses , any orders will be refunded
NeuroTrac® Myoplus is an all-in-one assessment and treatment tool.
It combines
- EMG biofeedback
- Electrical stimulation and
- ETS ( EMG triggered stimulation) modes. ETS allows you to set the threshold at which stimulation is triggered - meaning the patient has to contract to a pre-set threshold before the stimulation phase starts,
thus assisting in return to independent activation of the pelvic floor. - It can be used to assess electrical activity in the pelvic floor muscles, then switch to ETS or Stimulation mode for treatment - all in the one session.
NeuroTrac® Myoplus is a Portable Desk Top device, supplied with a desk top stand.
- EMG + Stimulation mode {Passive + Active}, EMG Triggered Stimulation, Muscle Stimulation and EMG Modes
- Designed to be used for a wide range of Physiotherapy, Sports and Rehabilitation applications including Continence and Stroke treatment.
- Custom programs enabling up to 5 phases to be set up in one program.
- Optional Multilingual PC Database Software
- Used as a learning tool with emphasis on improving treatment protocols using EMG and Neuromuscular Stimulation
- Locks programmes and records Patient Home Compliance on a daily basis.
- Portable, light and user friendly.
1. EMG
- Single Channel
- EMG Range: 0.2 to 2000 μV RMS (continuous)
- Sensitivity: 0.l μV RMS
- Accuracy: 4% of μV reading +/-0.3 μV at 200 Hz
- Selectable Bandpass filter - 3db Bandwidth,
- a. Wide: 18 Hz +/- 4 Hz to 370 Hz +/- 10% - Reading below 235 microvolts 10 Hz +/-3 Hz to 370 Hz +/- 10% - Readingabove 235 microvolts
- b. Narrow: 100 Hz +/- 5% to 370 Hz +/- 10%
- Notch filter: 50 Hz (Canada 60Hz) - 33 dbs (0. 1% accuracy)
- Common Mode Rejection Ratio: 130 dbs Minimum @ 50 Hz
- Battery: PP3 Alkaline
- Work / Rest periods: 2-99 seconds
- Number of Trials: 1-99
2. Neuromuscular Stimulation
- Single Channel
- Amplitude: 0-90 mA into 500 Ohm load - actual mA will tend to be less than indicated due to Electrode impedance: at 1000 Ohms load (Electrodes in poor condition) the maximum will be limited to 75 mA, at 1500 Ohms load the maximum will be limited to 50 mA.
- Type: Constant current, maximum output voltage 70 Volts +5 / -10 Volts
- Waveform: Symmetrical, rectangular, bi-phasic with net zero DC current
- Pulse width selection: 50-450 μS (2% accuracy)
- Pulse rate selection: 2-100 Hz (2% accuracy)
- Work / Rest periods: 2-99 seconds
- Time 1 - 99 minutes
- Ramp up time: 0.1 - 9.9 seconds
- Preset and user programmable treatment Programs
- Automatic output shut off with detection of open electrode above 0.5 mA
Specifications | 1. EMG
2. Neuromuscular Stimulation
Low battery indication at 7.4 volts +/- 0.2 volts and automatic shut off 4 minutes
|
---|
1. EMG
- Single Channel
- EMG Range: 0.2 to 2000 μV RMS (continuous)
- Sensitivity: 0.l μV RMS
- Accuracy: 4% of μV reading +/-0.3 μV at 200 Hz
- Selectable Bandpass filter - 3db Bandwidth,
- a. Wide: 18 Hz +/- 4 Hz to 370 Hz +/- 10% - Reading below 235 microvolts 10 Hz +/-3 Hz to 370 Hz +/- 10% - Readingabove 235 microvolts
- b. Narrow: 100 Hz +/- 5% to 370 Hz +/- 10%
- Notch filter: 50 Hz (Canada 60Hz) - 33 dbs (0. 1% accuracy)
- Common Mode Rejection Ratio: 130 dbs Minimum @ 50 Hz
- Battery: PP3 Alkaline
- Work / Rest periods: 2-99 seconds
- Number of Trials: 1-99
2. Neuromuscular Stimulation
- Single Channel
- Amplitude: 0-90 mA into 500 Ohm load - actual mA will tend to be less than indicated due to Electrode impedance: at 1000 Ohms load (Electrodes in poor condition) the maximum will be limited to 75 mA, at 1500 Ohms load the maximum will be limited to 50 mA.
- Type: Constant current, maximum output voltage 70 Volts +5 / -10 Volts
- Waveform: Symmetrical, rectangular, bi-phasic with net zero DC current
- Pulse width selection: 50-450 μS (2% accuracy)
- Pulse rate selection: 2-100 Hz (2% accuracy)
- Work / Rest periods: 2-99 seconds
- Time 1 - 99 minutes
- Ramp up time: 0.1 - 9.9 seconds
- Preset and user programmable treatment Programs
- Automatic output shut off with detection of open electrode above 0.5 mA
Low battery indication at 7.4 volts +/- 0.2 volts and automatic shut off 4 minutes
after pressing last key, unless infra-red is turned on.
Note: Replace battery immediately.
Environmental Conditions for use:
+10 to +30 degrees Centigrade. 0-90% Humidity.
Environmental conditions for storage & transport:
-10 to +50 degrees Centigrade. 0-90% Humidity.
Physical Dimensions: Length 147 mm, Width 70 mm, Depth 28.9 mm
Weight
MyoPlus device: 0.13 Kg (without battaery)
PP3 Alkaline battery: about 0.04 Kg
-
Everyway anal electrode$36.00